Requirements

  • Optimum personal, product and contamination protection
  • Handling of microbiological agents of classes S1, S2, S3 and S4
  • Integration of a liquid handling station for processing toxic and non-toxic samples

Tested

Certified to DIN EN 12469, NSF 49, Type A2
  • Solution

    The new generation of safety workbenches provides protection at the highest level. The integration of the BRAND Liquid Handling Station into a claire® series safety cabinet results in a unique combination that enables efficient processing of toxic and non-toxic samples.
    The claire® lh safety cabinet gives top priority to personal, product and contamination protection.

  • Functional principle

    In class II safety cabinets, a portion of the filtered exhaust air is blown down through the workspace in a laminar flow as an “air stamp”, ensuring product protection.

    Along the partially opened front window and together with the drawn-in room air an “air curtain” is created. This is returned to the filter or air recirculation through the air slots in the table segments, which protects the operator.
    Part of the exhaust air returns to the laboratory environment via the exhaust air filter, free of particles and microorganisms. This also minimises the risk of particles from
    inside the cabinet entering the laboratory environment.

  • Characteristics
    • High-End-Touch-Display
    • Optimal ergonomis
    • Liquid handling station
    • Intutive pipetting process
    • HEPA cartridge filter for low noise level
    • Ergonomic, especially quiet, bright working conditions, optimum legroom
    • modified three-piece worktop set
    • Dimmable LED lighting
  • Field of application
    • Pharmaceutical industry
    • Microbiology and cell biology
    • Medical technology
  • Additional products and options
    • Electrically height-adjustable base stand
    • UVC light
    • Utility supply connections and particle monitoring
    • Interfaces (e.g. USB)
    • FlexDuct outgoing air connection
    • Further options upon request

SKAN Pure Solutions

Technical specification

Technical Data (in mm)claire® lh 130Liquid Handling Station
Outer dimensions (W x D x H)1353 x 815 x 2008 – 2277594 x 485 x 530
Inside dimensions, working space (W x D x H)1257 x 600 x 640 – 700
Volume ranges1-Kanal Liquid Ends: 1 – 50 µl, 10 – 200 µl, 40 – 1000 µl 8-Kanal Liquid Ends: 1 – 50 µl, 20 – 300 µl, 40 – 1000 µl