Requirements
- Safe handling of active and toxic substances in powder form such as active pharmaceutical ingredients
- Sampling and product transfer
- Weighing in accordance with the latest USP requirements with analytical, precision and ultra-micro balances
Tested
Fulfils containment test, equivalent to robustness test SN EN 14175-3Weighing certificates according to USP chapter 41
OEL level <1 μg/m3 (according to exposure measurement)
Focus on the future
E-mode
Cost-saving operation
Motion sensor
The ‘e’ in e-mode stands equally for economic (economical) and ecological (ecological).
The e-mode ensures cost-saving and resource-conserving operation. This is achieved by reducing the air speed, whereby the safety function of the system is not impaired in any way and the environment remains demonstrably protected. Please refer to the ‘Containment test’ section. The e-mode is explicitly not a working mode. Motion sensors recognise whether work is currently being carried out on the system. In the operating settings, you can specify whether you want to switch to e-mode manually (via the touchscreen) or automatically (5, 10, 20 minutes after the system is used) to switch to e-mode. A blue light signal indicates to the user that the system is in energy-saving mode.
Downloads
Technical specification
Technical Data (in mm) | HFCpro 90 | HFCpro 120 | HFCpro 150 |
---|---|---|---|
Outer dimensions (W x D x H) D including armrest | 926 x 860 x 971 | 1226 x 860 x 971 | 1500 x 854 x 970 |
Usable work surface (W x D) (ceramic work surface) | 816 x 490 | 1116 x 490 | 1440 x 490 |
Air volume | ca. 375 m3/h | ca. 444 m3/h | ca. 554 m3/h |
Air velocity working mode | ca. 0,35 m/s | ca. 0,35 m/s | ca. 0,35 m/s |
Power consumption | max. 170 W | max. 200 W | max. 250 W |
Noise Level | 47,5 dB | 52 dB | – |