Sustainable
Safety workbench
claire® neo
The latest generation of safety cabinets for optimal safety
- Standard
- DIN EN 12469, DIN EN 12980
- Filter type
- HEPA 14
Simplifies routine cleanroom monitoring
Standard
21 CFR Part 11, GMP Annex 1
ISO class
ISO 5 or higher
During cleanroom qualification, the integrated assistant supports compliant performance. Simple programming of the measurement plans via area, location and other parameters – saves time and prevents operating errors.
Standardized workflows can be directly incorporated
Portable airborne particle counter for flexible applications across different cleanrooms and zones
A digital assistant supports the correct and documented execution of all measurements
Secure digital transmission of measurement values reduces errors and simplifies archiving
Password-protected access ensures that only authorized personnel can access the system
Designed for use according to EN ISO 14644 and GMP – reliably up to ISO Class 5 and beyond
All activities are recorded in a tamper-proof audit trail for maximum traceability
Support for Microsoft Active Directory allows centralized user management and easy IT integration
The air particle counter is ideal for accurately determining particle concentrations in air, compressed air, nitrogen, or noble gases. This allows companies to detect risks early, optimize process conditions, and ensure compliance with quality standards in sensitive production environments
The solution supports continuous classification and monitoring of cleanrooms according to EN ISO 14644 or GMP.
The particle counter enables the inspection of filter and exhaust ducts in critical systems. This allows early detection of malfunctions, optimization of maintenance intervals, and increased operational safety.
Measuring particle concentrations before and after HEPA filters provides reliable data on filter efficiency
A laser-based light scattering sensor is used to measure the number and size of airborne particles. With this technology, particles between 0.3 and 10 µm can be accurately detected.
The measurement data can be exported by USB stick, Ethernet or web interface. Alternatively, the data can also be printed directly with the integrated printer.
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0
28,3 LPM (1.0 CFM) ± 5 %
26.7 x 21 x 26
6.5
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0
28,3 LPM (1.0 CFM) ± 5 %
26.7 x 21 x 26
6.5
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0
50 LPM (1.77 CFM) ± 5 %
26.7 x 21 x 26
6.5
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0
50 LPM (1.77 CFM) ± 5 %
26.7 x 21 x 26
6.5
0.3 / 0.5 / 1.0 / 3.0 / 5.0 / 10.0
100 LPM (3.53 CFM) ± 5 %
26.7 x 21 x 26
6.9