The HIAC 9703+ particle counter is used for liquids in the pharmaceutical quality control

Tailored to your particle measurement needs, programmes can be individually configured. The PharmSpec software simplifies data management and allows secure access in different user levels.

HIAC at a glance

Characteristics

  • Integration of SOP’s

    Standardized workflows can be directly incorporated

  • Paperless data transfer

    Secure digital transmission of measurement values reduces errors and simplifies archiving

  • Compliant protection

    Password-protected access ensures that only authorized personnel can access the system

  • Complete documentation

    All activities are recorded in a tamper-proof audit trail for maximum traceability

  • Seamless integration

    Support for Microsoft Active Directory allows centralized user management and easy IT integration.

  • Compatibility

    with Windows 10, Windows 11 Pro and Enterprise

Requirements

Requirements that make your daily work safer and easier

  • Particle measurement in liquids

    The system enables precise determination of particles in ultrapure water, solvents, water for injection, as well as fuels and other liquids. This allows purity levels to be reliably verified and critical contaminants to be detected at an early stage.

  • Quality assurance and calibration

    Ideal for quality control applications where particle contamination must be clearly verified. Documented and traceable calibrations ensure measurement accuracy and reproducibility. This builds confidence during audits and meets the requirements of GMP and ISO standards.

Close up of the liquid particle counter 9703+
Functional principle

Precise particle measurement for a wide range of liquids

A laser sensor based on the light blocking principle is used to measure the number and size of particles. With this technology, particles between 1.3 and 600 µm can be accurately detected.
The particle counters are used for a variety of liquids, such as purified water, fuels, alcohols, hydrogen peroxide. The measurement data can be exported via PharmSpec software.

Liquid particle counter HIAC 9703+ close up touch panel

Technical specification

  • Particle size μm

    0.5 – 600

  • Flow rates ml/min.

    10 – 100

  • Outer dimensions (B x T x H) mm

    343 x 337 x 482

  • Weight

    ca. 10.7 kg