Skip to content
Laminar flow filter module Modulair
Back
Laminar flow

Filter module Modulair

Product protection on a grand scale

Application
Biotech & Life Sciences Laboratory & Research Food Industry Mechatronics & Medtech Hospital & Pharmacy
Protection category
Product protection
  • Standard

    EN ISO 14644-1

  • ISO-class

    ISO 5-8

The Modulair is intended for the creation of a clean area for isolated applications

The environment created by the filter modules protects the product from airborne particles and microorganisms. The integrated HEPA filter modules create a class 5–8 cleanroom zone in accordance with EN ISO 14644-1.

Modulair at a glance

Characteristics

  • Pure air quality

    HEPA filters class 14 as standard reliably remove even the smallest particles

  • Optimized airflow

    Vertical airflow ensures uniform distribution

  • Flexible solutions

    Individually configurable units that integrate seamlessly into existing production environments

  • Variable wall systems

    Available with louvres or fixed walls, adaptable to customer requirements

  • Low-noise EC fans

    Energy-efficient motors with reduced sound levels

  • LCD display with alarm function

    Intuitive display of all operating parameters

Requirements

Requirements that make your daily work safer and easier

  • Reliable product protection

    Sensitive products or sterile components remain protected throughout handling. This ensures product quality and extends shelf life.

  • Safe production environment

    A controlled production environment prevents airborne particles or microorganisms from entering the manufacturing process. This reduces contamination risks and supports compliance with regulatory requirements.

laminar flow-filter module Modulair close-up castors
Functional principle

Laminar airflow for maximum product protection

The ambient air is drawn in through a pre-filter and flows vertically into the working area through a high-performance HEPA filter. This laminar air flow is guided over the working area and flows out of the clean working area. It therefore protects the product or process from contamination.

Airflow scheme für Laminar Flow Modulair

Technical specification

  • Outer dimensions (W x D x H) w/o base stand in mm

    1280 x 695 x 900

  • Outer dimensions (W x D x H) with base stand in mm

    1280 x 695 x 2345

  • Inside dimensions, working space (W x D x H) in mm

    1200 x 635 x 665

  • Outer dimensions (W x D x H) w/o base stand in mm

    1890 x 1000 x 900

  • Outer dimensions (W x D x H) with base stand in mm

    1890 x 1000 x 2345

  • Inside dimensions, working space (W x D x H) in mm

    1810 x 940 x 665

  • Outer dimensions (W x D x H) w/o base stand in mm

    1595 x 1940 x 650 (Plenum)

  • Outer dimensions (W x D x H) with base stand in mm

    Individuell

  • Outer dimensions (W x D x H) w/o base stand in mm

    1940 x 2345 x 650 (Plenum)

  • Outer dimensions (W x D x H) with base stand in mm

    Individuell